User`s manual
This function appears in the MEASure and CONFigure commands as:
MEAS:TEMP? THER,type,(@channel_list)
CONF:TEMP THER,
type,(@channel_list)
Four-wire measurements are specified as:
TEMP FTHermistor,type
This function appears in the
MEASure and CONFigure commands as:
MEAS:TEMP? FTH,type[,(@channel_list)]
CONF:TEMP FTH,
type[,(@channel_list)]
Thermocouple
Measurements
Thermocouple measurements require the HP E1344A, E1347A, E1353A, or
HP E1476A multiplexers which are thermocouple compensated. The
thermocouple types supported are B, E, J, K, N14, N28, R, S, and T.
Thermocouple measurements are specified as:
TEMP TCouple,type
and appear in the
MEASure and CONFigure commands as:
MEAS:TEMP? TC,type,(@channel_list)
CONF:TEMP TC,
type,(@channel_list)
You can also measure the temperature of the reference thermistor on the
HP E1344A, E1347A, E1353A, or E1476A multiplexers as shown below:
MEAS:TEMP? THER,5000,(@nn93)
where nn is the multiplexer card number.
RTD Measurements The RTD types supported are 85 (alpha = 0.00385) and 92 (alpha = 0.00392).
RTD measurements can be made in either a 2-wire or 4-wire configuration.
Two-wire measurements require the scanning multimeter.
Two-wire RTD measurements are specified as:
TEMP RTD,type
The function appears in the
MEASure and CONFigure commands as:
MEAS:TEMP? RTD,type,(@channel_list)
CONF:TEMP RTD,
type,(@channel_list)
Chapter 4 Understanding the HP E1326B/E1411B Multimeter 89