User`s manual

Example Querying the Line Reference Frequency
CAL:LFR 50 !Reference frequency is 50 Hz.
CAL:LFR? !Query for reference frequency.
enter statement !Enter value into computer.
:ZERO:AUTO
CALibration:ZERO:AUTO <mode> enables or disables the autozero mode for DC
voltage and resistance measurements.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<mode>
boolean OFF | 0 | ON | 1 | ONCE none
Comments You can substitute decimal values for the OFF ("0") and ON ("1") parameters.
When autozero is ON, the multimeter makes a zero measurement (measurement
with input disabled) following every measured reading and subtracts the zero
measurement from the reading. This doubles the time required per reading.
When autozero is OFF, the multimeter makes one zero measurement and subtracts
this from all subsequent measurements. A new zero measurement is made
whenever the function is changed. Notice that the mode parameters
OFF and
ONCE have the same effect.
An autozero measurement is not made following a range change whether autozero
is
ON or OFF.
Autozero cannot be used when a 10 µs aperture time is specified.
The CONFigure and MEASure commands turn autozero ON.
*RST Condition: CAL:ZERO:AUTO ON
Example Turning Autozero OFF
CAL:ZERO:AUTO OFF !Disable autozero.
:ZERO:AUTO?
CALibration:ZERO:AUTO? returns a number to show whether the autozero mode
is enabled or disabled: "
1" = ON, "0" = OFF or ONCE.
The number is sent to the output buffer.