User`s manual

Measuring Temperature Using Thermocouples
(Agilent E1353A Module only)
The thermocouple terminal module automatically configures the Agilent
E1353A FET Multiplexer Module for thermocouple readings. The
multimeter measures the resistance of a thermistor on the terminal module
to provide a reference temperature. This reference temperature is used to
compensate for the change in temperature between the thermocouple
(desired measurement) and the terminal temperature (see ā€œCommentsā€ later
in this section).
Setup for
Measuring
Thermocouple
Temperature using
an External
Multimeter
Figure 3-7 shows how to set up channel 01 of a single module switchbox to
measure temperature using a thermocouple and an external multimeter.
Figure 3-7. Temperature Measurement Using a Thermocouple
Chapter 3 Using the Agilent E1351A/53A FET Multiplexer Modules 43