User`s manual

The logical addresses noted in Figures 2-1, 2-2, and 2-3 apply to modules
installed in an Agilent 75000 Series B Mainframe (Agilent Model Number
E1300B/E1301B) or in a mainframe with an Agilent E1405/1406 Command
Module. See the Agilent 75000 Series B Installation and Getting Started
Guide or the appropriate Agilent Command Module Manual for more
information on switchboxes and scanning voltmeter configurations, and
logical addressing. For uses in other systems or mainframes, see the
appropriate manuals.
Setting the
Address Switch
The logical address switch (LADDR) factory setting is 112. You may have
changed the switch setting during module installation. Valid address values are
from 1 to 255. Refer to the Agilent 75000 Series B System Installation and
Getting Started Guide or the appropriate Agilent Command Module Manual for
addressing information. Otherwise, refer to Figure 2-4 to reset the factory
setting.
Figure 2-3. Card Numbers for Scanning Voltmeter
Figure 2-4. Setting the Logical Address Switch
Chapter 2 Configuring the Agilent E1351A/53A FET Multiplexer Modules 19