Service manual

components.
After you remove an assembly from the multimeter, place the
assembly on a conductive surface to guard against ESD damage.
Do not stack assemblies.
Do not use pliers to remove a MOS or CMOS device from a
high-grip socket. Instead, use a small screwdriver to pry the
device up from one end. Slowly lift the device up, one pair of
pins at a time.
After you remove a MOS or CMOS device from an assembly,
place the device onto a pad of conductive foam or other suitable
holding material.
If a device requires soldering, be sure the assembly is placed on a
pad of conductive material. Also, be sure you, the pad, and the
soldering iron tip are grounded to the assembly. Apply as little
heat as possible when soldering.
When you replace a MOS or CMOS device, ground the foam to
the multimeter before removing the device from the foam.
Figure 8-2. Removing Binding Posts