Service manual

250 CLEAR SCREEN
260 GOTO Re_try
270 ELSE
280 PRINT "Adjustment complete with";Source(I);"Ohms source"
290 PRINT "Press Continue for the next adjustment"
300 PAUSE
310 CLEAR SCREEN
320 END IF
330 NEXT I
340 PRINT "Resistance adjustments completed"
350 END
Calibration
Errors
Table 5-1 summarizes calibration error numbers, titles and descriptions
for the Agilent E1326B multimeter. Note that an error is returned if the
adjustment (calibration) standard used is outside the calibration range
of the multimeter (between ± (0.5 full scale and full scale)).
Table 5-1. Agilent E1326B Multimeter Calibration Errors
Error Title Description Code*
0 No Error No error has occurred since last time the error code was read.
4 Reading
Overrun
The FIFO memory was still full at the time the new reading was
ready, or a new command was received while in the middle of
outputting a reading. TRIGGER ARM is disabled and the
multimeter waits for a new command.
U
6 Calibration
Error
An error occurred when computing a calibration constant,
probably due to an improper input or a defective Unit Under
Test (UUT). Calibration cycle aborted.
U/H
7 Checksum
Error
The non-volatile RAM contains a checksum error. The data is
assumed to be corrupted.
H
8 Invalid CAL
Request
Calibration requested for an invalid combination of multimeter
range and function.
U
9Bad CAL
Target
The target value for the calibration was outside the range of
± (0.5 full scale to full scale).
U
11 No Inguard
Response
No response from inguard (expected data and got nothing).
Timed out instead.
H
13 Linearity
CAL Error
An error has occurred during a linearity calibraton sequence.
Probably a hardware failure of the A/D inguard section.
H
5-10 Adjustments