AC Voltage
Adjustments
This procedure adjusts the Agilent E1326B AC voltage measurement
accuracy.
NOTE
The DC Voltage adjustment MUST be performed before the AC Voltage
adjustment.
Equipment
Setup
1. Connect the equipment as shown in Figure 5-2. You can connect
to the E1326B OR the E1326-80005 terminals, but not to both.
2. Set the Agilent E1326B as follows:
Reset Multimeter ......................................................*RST
Auto Zero .................................................................... ON
E1300B/E1301B
Figure 5-2. AC Voltage Adjustment Setup
Adjustments 5-5