Service manual

DC Voltage
Adjustments
This procedure adjusts Agilent E1326B DC voltage measurement
accuracy.
Equipment
Setup
1. Connect the equipment as shown in Figure 5-1. You can connect
to the E1326B OR the E1326-80005 terminals, but not to both.
WARNING
The DC Standard (Datron 4708, Option 10) can produce
dangerous voltages whch are present on the terminals. Do not
touch the front (or rear) panel terminals unless you are sure no
dangerous voltage is present.
2. Set the Agilent E1326B as follows:
Reset Multimeter ........................................................*RST
Auto Zero ......................................................................ON
Power Line Cycles (PLC) .................................................1
Line Freq Reference (CAL:LFR ) ............. 50 Hz or 60 Hz
E1300B/E1301B
Figure 5-1. DC Voltage Adjustment Setup
5-2 Adjustments