Service manual

Table 4-1. Performance Test Record for the Agilent E1326B Multimeter (Page 3 of 3)
Model ________________________________ Report No. ____________________________ Date_____________
1-Year Specifications
Test
No.
Test
Input
DMM
Range
Low
Limit
Measured
Reading
High
Limit
Meas
Uncert*
Test
Acc
Ratio
(TAR)**
DC Voltage (Zero Volts Input) (Values in Vdc)
4-1 0
0
0
0
0
0.113
0.91
7.27
58.1
300
-.000005
-.000015
-.00005
-.001
-.005
__________
__________
__________
__________
__________
+ .000005
+ .000015
+ .00005
+ .001
+ .005
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
DC Voltage (DCV Input) (Values in Vdc)
4-2 0.1
0.9
7.0
58.0
300.0
0.113
0.91
7.27
58.1
300
0.09996
0.89978
6.9986
57.98
299.9
__________
__________
__________
__________
__________
0.10004
0.90022
7.0014
58.02
300.1
.0000030
.0000046
.0000115
.0001680
.0011770
> 10:1
> 10:1
> 10:1
> 10:1
> 10:1
AC Voltage (60 Hz, 5 kHz, 10 kHz, 5 kHz) (Values in Vac)
4-3 0.07
0.07
0.07
300.00
0.0875
0.0875
0.0875
300
.0693
.0676
.0676
269.1
__________
__________
__________
__________
.0707
.0724
.0724
330.9
.000044
.000044
.000044
.038030
> 10:1
> 10:1
> 10:1
> 10:1
4-Wire Resistance (Values in Ohms)
4-4 1000
100000
1000000
2000
131000
1000000
999.6
99959
999590
__________
__________
__________
1000.4
100041
1000410
0.003
0.3
10.0
> 10:1
> 10:1
> 10:1
* Measurement Uncertainty of Datron 4708 source for 90 days since calibration and 23
o
C ± 1
o
C.
** TAR = multimeter accuracy/measurement uncertainty, shown as > 10:1.
Verification Tests 4-17