1 ! 4-Wire Ohms Performance Verification
2 !
10 OPTION BASE 1
20 DIM Range(3),Source(3),Read_meas(3),Limit(3),Ohms(3)
30 DATA 1861, 119156, 1048576
40 READ Range(*)
50 DATA 1000, 100000, 1000000
60 READ Source(* )
70 OUTPUT 70903;"*RST" Sets autozero on and PLC
1
80 OUTPUT 70903;"CAL:LFR 60" Set 60 Hz line ref freq
90 FOR I= 1 TO 3
100 PRINT "1. Set Resistance Standard to";Source(I);"Ohms"
110 PRINT "2. Measure ACTUAL Resistance Standard value (in
Ohms)"
120 INPUT " Enter ACTUAL Resistance Standard value (in Ohms)
",Ohms(I)
130 CLEAR SCREEN
140 OUTPUT 70903;"CONF:FRES ";Range(I) Set 4-wire ohms,
range
150 OUTPUT 70903;"READ?"
160 ENTER 70903;Read_meas(I)
170 IF I= 1 THEN Limit(I)= .0004*Ohms(I)+ 2.0E-2 2 kOhm
limits
180 IF I= 2 THEN Limit(I)= .0004*Ohms(I)+ 1. 131 kOhm
limits
190 IF I= 3 THEN Limit(I)= .0004*Ohms(I)+ 10. 1 MOhm
limits
200 NEXT I
210 PRINT " Measured Source Low Limit High Limit"
220 PRINT " Resistance Resistance (Ohms) (Ohms)"
230 PRINT
240 Format:IMAGE 7D.3D,6X,7D.3D,6X,7D.3D,6X,7D.3D
250 FOR I= 1 TO 3
260 PRINT USING Format;Read_meas(I),Ohms(I),
Ohms(I)-Limit(I),Ohms(I)+ Limit(I)
270 NEXT I
280 END
Verification Tests 4-13