Service manual
Test 4-4: Resistance Test (4-Wire Ohms)
This test verifies the 4-wire resistance accuracy of the 2kΩ, 131kΩ, and
1MΩ ranges.
NOTE
The DC Voltage performance test must be performed prior to the
Resistance Test to check the A/D accuracy on all ranges. If the DC
Voltage test has not been performed, resistance must be checked on all
ranges at 0 and at 50% of full scale.
Equipment Setup 1. Connect the equipment as shown in Figure 4-4. You can connect
to the E1326B OR to the E1326-80005, but not to both.
2. Set the Agilent E1326B multimeter as follows:
•
Reset Multimeter ....................................................*RST
• Auto Zero ..................................................................ON
• Power Line Cycles (PLC) .............................................1
• Line Freq Reference (CAL:LFR) .......... 50 Hz or 60 Hz
NOTE
*RST sets Auto Zero to ON and Power Line Cycles to 1.
E1300B/E1301B
Figure 4-4. Resistance Test Setup
Verification Tests 4-11