Service manual

WARNING
The AC Standard (Datron 4708, Option 20) can produce
dangerous voltages which are present on the terminals. Do not
touch the front (or rear) panel terminals unless you are sure no
dangerous voltage is present.
2. Set the Agilent E1326B multimeter as follows:
Reset Multimeter ..................................................*RST
Auto Zero ................................................................ON
Power Line Cycles (PLC) ...........................................1
Line Freq Reference (CAL:LFR) ........... 50Hz or 60Hz
NOTE
*RST sets Auto Zero to ON and Power Line Cycles to 1.
Test Procedure 1. Set the AC Standard (Datron 4708, Option 20) Output to 0.07
Vac at 60 Hz sine wave.
2. Set the Agilent E1326B range to 79.5 mV (87.5 mV with 10%
overrange) using CONF:VOLT:AC .07.
3. Measure the AC input voltage with READ? and verify the
results are within specified limits (at the range selected for1
PLC).
4. Repeat steps 1 through 3 using the following AC Standard
voltage and frequency settings, and Agilent E1326B ranges:
E1326B
Range
10%
overran
ge
AC Standard Output
Voltage
(Vac)
Frequen
cy (Hz)
79.5 mV
79.5 mV
300 V
87.5 mV
87.5 mV
N/A
0.07 V
0.07 V
300 V
5 kHz
10 kHz
5 kHz
5. Remove power and disconnect test equipment.