Service manual
Test 4-2: DC Voltage Test (DCV Input)
This test verifies DC Voltage accuracy on all five ranges with DC
voltage inputs.
Equipment Setup 1. Connect the equipment as shown in Figure 4-2. You can connect
to the E1326B OR the E1326-80005 terminals, but not to both.
WARNING
The DC Standard (Datron 4708, Option 10) can produce
dangerous voltages which are present on the terminals.
Do not touch the front (or rear) panel terminals unless
you are sure no dangerous voltage is present.
2. Set the Agilent E1326B multimeter as follows:
•
Reset Multimeter .................................................*RST
• Auto Zero ................................................................ON
• Power Line Cycles (PLC) ...........................................1
• Line Freq Reference (CAL:LFR) ............50Hz or 60Hz
NOTE
*RST sets Auto Zero to ON and Power Line Cycles to 1.
E1300B/E1301B
Figure 4-2. DC Voltage (DCV Input) Setup
4-6 Verification Tests