Service manual
Recommended
Test Equipment
Table 1-1 lists the test equipment recommended for testing, adjusting
and servicing the multimeter. Essential requirements for each piece of
test equipment are described in the Requirements column.
Table 1-1. Recommended Test Equipment
Instrument Requirements Recommended Model Use*
Controller, GPIB GPIB compatibility as defined by IEEE
Standard 488-1978 and the identical
ANSI Standard MC1.1: SH1, AH1, T2,
TEO, L2, LE0, SR0, RL0, PP0, DC0, DT0,
and Cl, 2, 3, 4, 5
HP Series 300
or
IBM compatible PC with
BASIC
A,O,F,
P,T
Mainframe Compatible with multimeter Agilent E1300B, E1301B,
E1302A, or E1401B/T,
E1421A (requires E1405A/B
or E1406A)
A,O,F,
P,T
AC Standard Voltage Range 0.1 V to 300 V Datron 4708 with Option 20 A,P
DC Standard Voltage Range 0.07 V to 300 V Datron 4708 with Option 10 A,P
Resistance Standard
Values 1 kΩ to 1 MΩ
Datron 4708 with Option 30 A,P
Digital Multimeter General Purpose Voltage and Resistance Agilent 3458 T
*A = Adjustments, F = Functional Verification, M = Preventive Maintenance, O = Operation Verification Tests,
P = Performance Verification Tests, T = Troubleshooting
E1300B/E1301B
Figure 1-2. Agilent E1326B Field Installation Kits
General Information 1-5