Mainframes User's Manual
Agilent E1332A 4-Channel Counter/Totalizer Menu
Menu Levels and Content
Level 1 Level 2 Level 3 Level 4 Level 5 User Entry Command( s) Used Description
COUNTER MONITOR CHAN1 DISP:MON:CHAN 1;STAT ON Monitor instrument operations on channel 1
CHAN2 DISP:MON:CHAN 2;STAT ON Monitor instrument operations on channel 2
CHAN3 DISP:MON:CHAN 3;STAT ON Monitor instrument operations on channel 3
CHAN4 DISP:MON:CHAN 4;STAT ON Monitor instrument operations on channel 4
AUTO DISP:MON:CHAN AUTO;STAT ON Monitor instrument operations on active channel
INPUT LEVEL CHAN1&2 voltage † SENS1:EVEN:LEV < value> Set level trigger voltage for channels 1 & 2
CHAN3&4 voltage † SENS3:EVEN:LEV < value> Set level trigger voltage for channels 3 & 4
SLOPE CH AN1 POS SENS1: EVEN: SLOP POS Posi t i ve l evel t r i gger sl ope f or chan n el 1
NEG SENS1:EVEN:SLOP NEG Negative level trigger slope for channel 1
CHAN2 POS SENS2: EVEN:SLOP POS Positi ve level t ri gger slope for channel 2
NEG SENS2:EVEN:SLOP NEG Negative level trigger slope for channel 2
CHAN3 POS SENS3: EVEN:SLOP POS Positi ve level t ri gger slope for channel 3
NEG SENS3:EVEN:SLOP NEG Negative level trigger slope for channel 3
CHAN4 POS SENS4: EVEN:SLOP POS Positi ve level t ri gger slope for channel 4
NEG SENS4:EVEN:SLOP NEG Negative level trigger slope for channel 4
ISOLATE ON INP:ISOL ON Input isolation on
OFF INP:ISOL OFF Input isolation off
FILTER ON INP:FILT ON Input filter on
OFF INP:FILT OFF Input filter off
FREQ frequency ‡ INP:FILT:FREQ < value> Set input filter frequency
FREQ CHAN1 TRIG:SOUR IMM;:MEAS1:FREQ? Frequency measurement on channel 1
CHAN3 TRIG:SOUR IMM;:MEAS3:FREQ? Frequency measurement on channel 3
PERIOD CHAN1 TRIG:SOUR IMM;:MEAS1:PER? Period measurement on channel 1
CHAN3 TRIG:SOUR IMM;:MEAS3:PER? Period measurement on channel 3
(continued on following page)
2-24 Using the Display Terminal Interface