User`s guide
Contents ▲ 306 ▼ Index
NOTE
With bioanalyzer model G2938A only diagnostic tests in combination with the
electrode cartridge can be performed.
Test C hi ps
Depending on your bioanalyzer setup (electrophoresis or flow cytometry), different test
chips are required to run some of the diagnostics tests. Test chip kits are part of the
bioanalyzer electrophoresis set (G2947CA) and flow cytometry set (G2948CA):
Test Chip Kit for Electrophoresis Assays (reorder no. G2938-68100)
System Leak Test Checks if the bioanalyzer is able to maintain a vacuum.
Produces a test pressure of -100 mbar and monitors for
changes. Cell Autofocus test chip required.
Cell Autofocus Test Checks that the optical system of the bioanalyzer is correctly
calibrated. Cell Autofocus test chip required.
Test Chip Comment Quantity
Autofocus Test Chip Values for fluorescence and offset are printed on
the chip. Can be used multiple times.
1
Diagnostics Test Purpose