User`s guide

Contents 303 Index
Running Instrument Diagnostics
2100 expert provides several tests to check proper functioning of the bioanalyzer
hardware. You should perform the tests on a regular basis, or if incorrect measurements
occur.
You can test the following:
Generic bioanalyzer tests, which can be run with both types of cartridges (electrode or
pressure cartridge)
Bioanalyzer in combination with electrode cartridge (electrophoresis setup)
Bioanalyzer in combination with pressure cartridge (flow cytometry setup)
Generic Bioanalyzer Tests
Diagnostics Test Purpose
Electronics Test Verifies proper functioning of all electronic boards in the
bioanalyzer.
Lid Sensor Test Verifies proper operation of the lid sensor, ensuring that the
laser and LED are off when the lid is open.
Stepper Motor Test Checks for proper movement of the stepper motor.