Technical data
Cable Total Jitter Tests 22
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 199
9 In order to perform the equalization, you have the option to use your
own coefficient file. To do so, at the Configure page, under the Cable
Equalizer option, Select “User Defined File” from the drop down menu.
Test Plan Check Mode Turn on test plan check mode to simulate the actual test plan run flow
without actual tests being run.
SSC Acquisitions Number of SSC cycle captured for SSC related tests. Maximum
number is 25.
Low Bit Rate SSC
Smoothing Points
Sets the number of smoothing points for SSC low pass filter when bit
rate is 1.62 Gbps.
High Bit Rate SSC
Smoothing Points
Sets the number of smoothing points for SSC low pass filter when bit
rate is 2.70 Gbps.
Memory Depth Sets the memory depth for acquisition
Cable Test
Cable Equalization Selects the Type of Equalizer of the cable test.
Cable Construction Selects the Cable Construction used.
Cable Mask Selects the type of mask to use for the eye test.
Cable Mask Movement This field contains 3 options. (1) Find Passing Mode will automatically
search +/-0.5UI horizontally until no violation occurs, (2) Fixed Mask
will not be moving, it only report Pass or Fail upon test, (3) Find Biggest
Margin will search +/-0.5UI horizontally to find the maximum margin
of non-violation mask.
Jitter Separation Edges Sets the number of edges measured for the jitter separation test.
Bit Error Rate Sets the bit error rate for the RJ/DJ measurements.
Table 27 Test Configuration Options
Configuration Option Description