Technical data

130 U7232A DisplayPort Compliance Testing Notes
14 Calibrating the Infiniium Oscilloscope and Probe
Required Equipment for Calibration
To calibrate the Infiniium oscilloscope in preparation for running the
DisplayPort automated tests, you need the following equipment:
Keyboard, qty = 1, (provided with the Agilent Infiniium oscilloscope).
Mouse, qty = 1, (provided with the Agilent Infiniium oscilloscope).
Precision 3.5 mm BNC to SMA male adapter, Agilent p/n 54855- 67604,
qty = 2 (provided with the Agilent Infiniium oscilloscope).
Calibration cable (provided with the Agilent Infiniium oscilloscopes).
E2655A/B probe de- skew fixture.
82
damping resistors (01130- 81506) for the Socketed Differential
Probe Head.
Figure 48 below shows a drawing of the above connector items.
Figure 48 Accessories Provided with the Agilent Infiniium Oscilloscope
Precision 3.5 mm Adaptors (2)
BNC Shorting Cap
(Used to calibrate
54854A and 54855A
oscilloscopes)
BNC Calibration Cable
(Used to calibrate 54854A oscilloscope)
Calibration Cable
(Used to calibrate
54855A and 80000
Series
oscilloscopes)