Table 2- 8 Data log menu options
Option Available settings Description
TRIGGER
DELAY
0 to 3600 seconds Delay time between a trigger is initiated and first reading is taken by the data log
function. The smallest delay time resolution is 100 µs.
SAMPLE
INTERVAL
1 to 3600 seconds Delay time between subsequent readings. The smallest resolution is 100 µs. The
minimum interval is configuration dependent and may be lower than 1 s.
SAMPLE
COUNT
1 to 5000 (single display
measurement)
1 to 2500 (dual display
measurement)
Total readings that need to be logged. It can be configured from 1 to 5000
readings for single display measurement (upgrade to 50,000 with option
3445MEMU) and 2500 readings for dual display measurement (upgrade to
25,000 with option 3445MEMU).
TRIGGER
COUNT
1 to 5000 (single display
measurement)
1 to 2500 (dual display
measurement)
Total triggers that will be received. It can be configured from 1 to 5000 triggers
for single display measurement (upgrade to 50,000 with option 3445MEMU) and
2500 triggers for dual display measurement (upgrade to 25,000 with option
3445MEMU).
ENABLE
LOG
- To start the log function. During logging, all keys are locked. To stop the log
function, press any key except TRIG key follow by SHIFT key.