Technical data

8 Characteristics and Specifications
Digital Multimeter Specifications
252 U3606A User’s and Service Guide
Response time: Approximately 1 s for 100 μF and ranges below 100 μF
Charge and discharge
voltage:
5 V
pp
(approximately from +3 V to –2 V)
Frequency
Measurement method: Reciprocal counting technique
Signal level: 0.2 V to 1.4 V
Input protection:
Voltage path: 1000 V
rms
on all ranges, < 0.3 A short circuit
Current path: Protected with a 3.15 A/500 V, FF fuse
Maximum display counts (excluding frequency)
5½ digit: 120,000
4½ digit: 12,000
Measurement noise Rejection
CMRR (common mode
rejection ratio) for 1 kΩ
unbalanced in LO lead
DC: 140 dB
AC: 70 dB
NMRR (Normal Mode
Rejection Ratio)
60 Hz ± 0.1%
5½ digits: 65 dB
4½ digits: 0 dB
50 Hz ± 0.1%
5½ digits: 55 dB
4½ digits: 0 dB