196 U3606A User’s and Service Guide
Valid gain adjustment input values
Gain adjustment can be accomplished using the following input values.
Table 6 - 2 Valid gain adjustment input values
Function Range
Valid amplitude
input values
Page
DC voltage
100 mV
1 V
–1 V
10 V
100 V
1000 V
0.9 to 1.1 × full scale page 197
AC voltage
10 mV
100 mV
1 V
10 V
100 V
750 V
0.9 to 1.1 × full scale page 199
Frequency Autorange/1 kHz
Input ¥ 100 mV
rms
,
900 Hz to 1100 Hz
page 200
Resistance
(2-wire)
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
0.9 to 1.1 × full scale page 201
DC current
10 mA
100 mA
1000 mA
0.9 to 1.1 × full scale page 203