Technical data
System Related Operation 4
Utility Menu Summary
U3606A User’s and Service Guide 123
5 Press Shift > Save to save or press Shift > Exit to exit the mode without
saving.
6 Press Shift > Exit to exit the utility menu.
Configuring the ramp signal parameters
The following procedure shows you how to configure the ramp signal
parameters.
1 To access the utility menu, press Shift > Utility.
2 The first utility menu item (Error) will be shown in the upper
secondary display.
3 Press or until the menu item “rAMP” is shown in the upper
secondary display.
Table 4 - 3 Ramp signal parameters
Ramp signal Output
Items Constant Voltage Constant Current
Range S1S2S1S2
Amplitude end position
[1]
[1] Amplitude start position is fixed at 0 (V or A) by default.
0 V to 30.000 V 0 V to 8.400 V 0 A to 1.0000 A 0 A to 3.0000 A
Number of steps 1 step to 10000 steps
Dwelling time Fastest of the instrument output capability
[2]
[2] Typically ~300 ms per step.