Technical data
4 System Related Operation
Utility Menu Summary
122 U3606A User’s and Service Guide
Setting the output protection state
The OVP and OCP protection features depend on the output protection
state function. The output protection state must be enabled for the OVP
and OCP protection features to be active in the circuit. Disabling the
output protection state will deactivate both the OVP and OCP protection
features even if a trip level is set for either the OVP or OCP. See
Chapter 3, “Protection Functions,” starting on page 78 for more
information.
The following steps shows you how to enable or disable the output source
protection.
1 Press Shift > Utility to access the utility menu.
2 The first utility menu item (Error) will be shown in the upper
secondary display.
3 Press or until the menu item “triP Sby” is shown in the
secondary display as shown in the figure below.
4 Press or to switch between “YES” and “no”. Select “YES” to
enable the output protection state or “no” to disable the output
protection state.
NOTE
The output protection state does not affect the overvoltage and
overcurrent limits.
CAUTION
Disabling the output protection state may result in equipment damage
if an overvoltage or overcurrent condition occurs. It is recommended to
have the output protection state enabled.