Technical data

4 System Related Operation
Utility Menu Summary
112 U3606A User’s and Service Guide
IOb
GPib U-CdC U-tMC
Select “GPib”, “U-tMC”, or “U-CdC” as the desired remote
interface connection.
“USB-tMC” simulates the USB interface according to
USB-TMC standard.
“USB-CdC” is used to simulate the communication port.
See “Connecting to a remote interface” on page 118 and
“Remote Operation” on page 134 for more information.
NN
Set the GPIB address from 1 to 30.
See “Configuring and connecting the GPIB interface” on
page 135 for more information.
SELF.t no YES
If “YES” is selected, the instrument exits the utility menu
and immediately executes the self-test.
After the self-test completes, the instrument returns to
normal operation.
See “Performing a self-test” on page 120 for more
information.
db.rEF NNNN Ω
Set dB reference impedance value from 1 Ω to 9999 Ω.
See “Selecting a dBm reference resistance value” on
page 121 for more information.
triP Sby YES no
Select “YES” to enable the output protection or “no” to
disable the output protection.
The factory default for the output protection state is “YES”.
See “Setting the output protection state” on page 122 for
more information.
rAMP
CV CC
Press Voltage to select ramp setup for CV.
Press Current to select ramp setup for CC.
S1/S2 S1/S2
Press Shift > Range to toggle between range S1 (30 V/1 A) or
S2 (8 V/3 A)
NN.NNN V N.NNNN A NNNNN S
The increment of each step will be the amplitude end
position (NN.NNN V/N.NNNN A) divided by the number of
steps (NNNNN).
The ramp dwelling time will be fastest of output capability
(typically ~300 ms per step).
See “Configuring the ramp signal parameters” on page 123
for more information.
Table 4 - 2 Utility menu item descriptions (continued)
Item Available settings Description