Technical data
DC Power Supply Operation 3
Protection Functions
U3606A User’s and Service Guide 87
NOTE
• The OC limit feature is always enabled by default whenever you select
constant voltage mode. You cannot disable the OC limit feature.
• If the OC value is set to a greater value than the OCP value, the OCP
value will be adjusted to equal the OC value.
• If the OC value is set to zero, this will result in the output voltage being
dropped to zero for limiting.
• The OC value is not applicable for square-wave output.
• The OC value is limited by the range selected. Press Shift > Range to
select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).