User manual

If P1Control and/or P2Control are enabled for the Add/Subtract mode then the data will be added if the
signal, or the or of both signals, is in the high state. The same rule holds if they are used for trigger enable.
The P1Control/P2Control "average (out)" signal goes high after the first trigger is accepted for an average
and drops back down when the last trigger's acquition is complete.
Example
long channelNbr = 0, dither = 8;
AcqrsD1_configAvgConfigInt32(ID, channelNbr, "DitherRange", dither);
This function sets the dithering range to ± 8 LSB’s.
Note that this function takes value of the parameter to be set, not the the address.
LabWindowsCVI/Visual C++ Representation
ViStatus status = AcqrsD1_configAvgConfigInt32(ViSession instrumentID,
ViInt32 channelNbr, ViString parameterString,
ViInt32 value);
LabVIEW Representation
Please use the Acqiris Dx.lvlib: (or Aq Dx) Extended Configure Averager.vi
described in
AcqrsD1_configAvgConfig.
Visual Basic Representation
ConfigAvgConfigInt32 (ByVal instrumentID As Long, _
ByVal channelNbr As Long, _
ByVal parameterString As String, _
ByVal value As Long) As Long
Visual Basic .NET Representation
AcqrsD1_configAvgConfigInt32 (ByVal instrumentID As Int32, _
ByVal channelNbr As Int32, _
ByVal parameterString As String, _
ByVal value As Int32) As Int32
MATLAB MEX Representation
[status]= AqD1_configAvgConfigInt32(instrumentID, channel, parameterString,
value)
Programmer’s Reference Manual Page 67 of 210