User manual

The channelNbr is used to designate the channel number for those parameters whose values can be different
for the two channels of an AP240/AP235 in dual-channel mode. These parameters are indicated in bold in
the list above.
The granularity for "NbrSamples","StartDelay", and "StopDelay" is 16 for the AP100/AP101 and the
AP240/AP235 in Dual-Channel mode and 32 for the AP200/AP201 and the AP240/AP235 in Single-
Channel mode.
If P1Control and/or P2Control are enabled for the Add/Subtract mode then the data will be added if the
signal, or the or of both signals, is in the high state. The same rule holds if they are used for trigger enable.
The P1Control/P2Control "average (out)" signal goes high after the first trigger is accepted for an average
and drops back down when the last trigger's acquition is complete.
Example
long channelNbr = 0, dither = 8;
AcqrsD1_configAvgConfig(ID, channelNbr, "DitherRange", &dither);
This function sets the dithering range to ± 8 LSB’s.
Note that this function takes the address, not the value of the parameter to be set.
Programmer’s Reference Manual Page 62 of 210