Appendix Compensation
3. Electrical Length Compensation
In a single-port (Two-Terminal) impedance measurement at higher frequencies (RF region or higher), the wavelengths of
the electrical signal are so short that the length of the signal transmission line including the test fixture and cables can
cause an undesirable phase shift error. The phase shift error is corrected by the electrical length compensation, which
should be performed in combination with the open/short compensation.
The following measurement instruments are equipped with electrical length compensation function.
*denotestheinstrumentisobsolete.
The electrical length is specified for the following test fixtures:
Electrical length of a test fixture Residual parameters in the measuring circuit
Model Electrical Length
16092A 3.4 mm
16192A 11.0 mm
16194A 50.0 mm
16196A 26.2 mm
16196B 26.9 mm
16196C 27.1 mm
16196D 27.3 mm
16197A 14.0 mm
Model Compensation Function
4287A* Electrical Length Compensation/Port Extension
4291B* Electrical Length Compensation/Port Extension
4294A + 42942A Electrical Length Compensation/Port Extension
E5061B-3L5 + Option 005 + 16201A Electrical Length Compensation/Port Extension
E4982A Electrical Length Compensation/Port Extension
E4991A Electrical Length Compensation/Port Extension