Error Compensation
Agilent measurement instruments incorporate one of the following four types of error compensation functions to eliminate
residual impedance effects in test fixtures:
1. Open/short Compensation
For a simple measurement system, represented by the equivalent circuit model shown below, residual impedance val-
ues Rs and Ls and admittance values Co and Go can be corrected by:
(i) Measuring open condition for the test fixture’s admittance.
(ii) Measuring short condition for the test fixture’s impedance
(iii) Measuring the test sample, then subtracting the admittance
and impedance.
The above procedure is performed internally by the open/short
compensation. When the test fixture is directly connected to the
measurement instrument, the open/short compensation suffi-
ciently corrects the measurement error.
2. Open/short/load Compensation
When the measurement system is too complicated to be repre-
sented as the above equivalent circuit model, the open/short compensation cannot completely compensate for the
residual impedance. In this case, the open/short/load compensation is used rather than the open/short compensation.
The open/short/load compensation is particularly effective when, for example, the 16065A is used with the measure-
ment instrument.
The following table lists measurement instruments and available compensation functions.
–: N/A
●: Available
* denotestheinstrumentisobsolete.
Note: For more details of the compensation functions, refer to the instruction manuals supplied with
the measurement instrument.
Appendix Compensation
Test Fixture
Instrument
DUT
Model Open compensation Short compensation Load compensation Load's input parameters
4263B ● ● ● Z-Q, R-X, Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q,
Cs-Rs, Ls-D, Ls-Q, Ls-Rs
4268A* ● ● ● Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs
4279A* ● ● ● Cp-D, Cp-G
4284A* ● ● ● same as measurement parameters
4285A ● ● ● same as measurement parameters
4287A* ● ● – –
4288A* ● ● ● Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-Q, Cs-Rs
4291B* ● ● ● Rs-Ls
4294A ● ● ● Rs-Ls
4338B* – ● – –
4339B* ● – – –
4349B* ● – – –
E4980A ● ● ● same as measurement parameters
E4981A ● ● ● Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs
E4982A ● ● – –
E4991A ● ● – –