Specifications

55
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
16118A Tweezers Test Fixture
Terminal Connector:
* InterlockConnectorenablesanddisables
the application of source voltage from the
measurement instrument.
DUT Connection: 2-Terminal (with triaxial cable)
Cable Length (approx.): 1 m
Weight (approx.): 410 g
55
DC (High resistance) SMD
Description: The 16118A is designed to operate specifically with
4339B. The tweezer’s contacts, makes it easy to hold SMD.
Electrical noise effects are reduced by the employment of shielded-
cables. A built-in interlocking circuit enables safe high-voltage mea-
surements.
Applicable Instruments: 4339B*
*denotestheinstrumentisobsolete.
Frequency: DC
Maximum Voltage: 100 V
Maximum Current: 0.87 mA
Resistance Measurement Range: 1 x 10
7
to 1 x 10
11
Operating Temperature: 0°C to 55°C
DUT Size: See figure below
Furnished Accessories:
Compensation and Measurement: Open compensation is recom-
mended before measurement. Separate the electrodes from each
other. The separation should be equivalent in size to the DUT’s
width. After performing open compensation, the DUT is sandwiched
by the tweezers’ contacts and is measured.
4339B with 16118A
DUT Dimensions
Type Connector
Input Triaxial (special screw-type)
Output High Voltage BNC (special type)
Control Interlockconnector*
Description P/N Qty.
Operation and Service Manual 16118-90010 1