Up to 3 GHz (7 mm) Material
Placing the MUT on the Test Fixture as follows:
Furnished Accessories:
Compensation and Measurement: Short compensation is required
before measurement. Short compensation is performed by only
inserting the MUT holder into the test fixture. After performing short
compensation, the MUT is inserted into the fixture as shown below.
Remove the cap of the fixture Place a MUT holder only in the fixture Replace the cap by screwing tightly
Remove the cap of the fixture Place a MUT onto the MUT holder and insert it
into the fixture.
Replace the cap by screwing tightly
Short Compensation
E4991A with 16454A
Description P/N Qty.
Fixture Holder 16454-00601 1
Tweezers 8710-2081 1
Screw, Hex Recess 0515-1050 1
Holder A 16454-25002 1
Holder B 16454-25001 1
Holder C (Without hole) 16454-25003 1
Holder D (Without hole) 16454-25004 1
Holder Case 1540-0622 1
Hex Key (for replacing fixtures) 8710-1181 1
Carrying Case 16454-60101 1
Operation and Service Manual 16454-90020 1