Specifications

43
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
4343
Up to 3 GHz (7 mm) SMD
16196U Maintenance Kit
Opt. 16196U-010: Upper electrode, 5 piece set (common to
16196A/B/C models)
Opt. 16196U-300: Short plate for 0201 (inch)/0603 (mm) size, 5
piece set (for 16196C)
Opt. 16196U-310: Lower electrode, 5 piece set (for 16196C)
Compensation and Measurement: Open and short compensations
are recommended in combination with the electrical length compen-
sation before measurement. The fixture’s electrical length must be
entered into the electrical length compensation function of the mea-
surement instrument first. Next, open compensation is performed by
placing the furnished open plate on top of the insulator assembly.
Short compensation is performed by placing the furnished shorting
plate on top of the insulator assembly. After performing open and
short compensations in combination with the electrical length com-
pensation, the DUT is inserted into the test fixture. Once the mea-
surement of the DUT is complete, remove the DUT from the fixture,
by using the furnished push ring. Refer to the 16196A figures to see
how compensation and measurement is performed.