Up to 3 GHz (7 mm) SMD
Options:
16194A-010: Add EIA/EIAJ industrial standard sizedshorting bar set
16194A-701: Add general sized shorting bar set
Compensation and Measurement: Before beginning the measure-
ment, the appropriate device holder (for a SMD or lead component)
must be prepared with the text fixture. The following figure shows
how the device holder is exchanged to match the device type.
The next step is to perform open and short compensations in combi-
nation with the electrical length compensation. When measuring
above 500 MHz, load compensation is also recommended. The fix-
ture’s electrical length must be entered into the electrical length
compensation function of the measurement instrument first. Then
open compensation is performed by separating the high and the low
electrodes from each other. The separation should be equivalent in
size to the DUT’s width. Short compensation is performed by using
the option 16194A-010/701 shorting bar set. Load compensation is
performed by using the furnished 50 SMD chip resistor. After per-
forming open, short, and load compensations in combination with
the electrical length compensation, the DUT is inserted into the test
fixture. The following figures show how measurement is performed.
Exchanging the device holder
1. Remove the ground plate
2. When measuring SMD, attach the
knob on the device holder.
3. Select the device holder suitable for
the device type. Loosen its knob and
insert into the arm.
4. Set the ground plate.
Placing the device