Specifications

34
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
3434
Up to 3 GHz (7 mm) SMD
Options:
16192A-010: Add EIA/EIAJ industrial standard sized shorting bar
set
16192A-701: Add general sized shorting bar set
16192A-710: Add the magnifying lens and tweezers
Compensation and Measurement: Open and short compensations
are recommended in combination with the electrical length compen-
sation before measurement. The fixture’s electrical length must be
entered into the electrical length compensation function of the mea-
surement instrument first. Then open compensation is performed by
separating the high and the low electrodes from each other. The
separation should be equivalent in size to the DUT’s width. Short
compensation is performed by using option 16192A-010/701 short-
ing bar set. After performing open and short compensations in com-
bination with the electrical length compensation, the DUT is inserted
into the test fixture. The following figures show how compensation
and measurement is performed.
Test fixture overview
Open/Short compensation
For more details, please refer
to 16192A operation manual.