Specifications

24
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
24
Up to 110 MHz (4-Terminal Pair) DC Bias Accessories
16065A 200Vdc External Voltage
Bias Fixture
Terminal Connector: 4-Terminal Pair, BNC
DUT Connection: 4-Terminal
External Bias Input connector: High Voltage BNC(f)
Dimensions (approx.): 180(W) x 120(H) x 200(D) [mm]
Cable Length (approx.): 40 cm
Weight (approx.): 1500 g
High Voltage BNC(f) connector for external bias input
BNC(f) connector for voltage monitor output
LCR meter with 16065A
16065C 40Vdc External
Voltage Bias Adapter
Terminal Connector: 4-Terminal Pair, BNC
External Bias Input connector: BNC(f)
Dimensions (approx.): 160(W) x 50(H) x 150(D) [mm]
Cable Length (approx.): 210 mm
Weight (approx.): 450 g
LCR meter with 16065C
Description: This test fixture makes it possible to measure a DUT
with up to ±200 V DC bias. The same modules of 16047A can be
used to allow measurements of axial/radial lead components.
Applicable Instruments:4263B4268A*4279A*4288A*E4981A
4284A*4285A4294AE4980A**
*denotestheinstrumentisobsolete.
**applicableinalimitedreuencyrange.
Frequency: 50 Hz to 2 MHz
Maximum DC Bias: ±200 V DC max. /15 V peak AC max.
Blocking Capacitor of 5.6 µF is connected in with the Hc terminal.
Operating Temperature: 0°C to 55°C
DUT Size: See the 16047A figure with module sizes.
Furnished Accessories:
Compensation and Measurement: Open, short and load compensa-
tions are recommended before measurement. Short compensation is
performed by shorting the contacts of the test fixture with a shorting
plate as described for the 16047A. Load compensation is performed by
inserting a known standard device. After performing open, short and
load compensations, the DUT is connected to the test fixture.
Description: This adapter is designed to operate specifically with the
4263B, 4268A*, 4288A* and the E4981A. By connecting an external
DC voltage source to this adapter, a bias voltage of up to ±40 V can
be supplied to a DUT. The DUT can be inserted by connecting any
direct attachment 4-Terminal Pair test fixture to the adapter.
Applicable Instruments:4263B4268A*4288A*E4981A
Frequency: 100 Hz to 1 MHz
Maximum DC Bias: ±42 V peak max. (AC+DC)
Blocking Capacitor of 100 µF is connected in series with the Hc ter-
minal.
Operating Temperature: 0 °C to 55 °C
Applicable Fixtures: 16034E/G/H, 16044A,
16047A/D
*
/E,
16048A/B*/D/E16089A/B/C/D/E
*
*denotestheinstrumentisobsolete.
Furnished Accessories:
Compensation and Measurement: Open and short compensations are
recommended before measurement. Short compensation is performed
by shorting the contacts of the test fixture that is in use. After perform-
ing open and short compensations, the DUT is connected to the test
fixture.
Description P/N Qty.
Module For Axial Lead 16061-70022 1
Module For Radial Lead mounting on fixture 16061-70021 1
Module For Short Radial Lead 16047-65001 1
Shorting Bar 16047-00640 1
Operating and Service Manual 16065-90011 1
Description P/N Qty.
Operating and Service Manual 16065-90020 1