Specifications

11
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
11
16060A Transformer Test Fixture
Terminal Connector: 4-Terminal Pair, BNC
DUT Connection:
2-Terminal for L measurement
3-Terminal for N, M measurement
See figure below for more information.
Dimensions (approx.):
90 (W) x 35 (H) x 90 (D) [mm]
Cable Length (approx.): 25cm
Weight (approx.): 300 g
Additional Error: The additional error is negligible when compared to the
instrument’s accuracy.
11
Up to 110 MHz (4-Terminal Pair) Lead Components
Description: This test fixture provides a convenient means of mea-
suring a transformer’s self-inductance, mutual inductance, turns-
ratio, and DC resistance in the frequency range of DC to 100 kHz, as
appropriate for each measurement.
Applicable Instruments: 4263B with Option 4263B-001
Frequency: DC to 100 kHz
Operating Temperature: 0°C to 55°C
DUT Size: The lead wire of the transformer should not have a diame-
ter greater than 4 mm, otherwise the alligator clip will not be able to
clamp onto it properly.
Furnished Accessories:
Compensation and Measurement: Open compensation is
recommended before measurement. Open compensation is performed
by connecting the alligator clips of “A” and “B” terminals together
and separating them from the likewise connected alligator clips of
the COMMON terminals. After performing open compensation, the
transformer is connected to the test fixture. The “A” and “B”
terminals are connected to the high terminals of the transformer.
The COMMON terminals are connected to the low terminals of the
transformer. The following figures show how compensation and
measurement are performed.
Open compensation Connecting a transformer
4263B with 16060A
Description P/N Qty.
Test Leads (black), Alligator clip to BNC(m) 16060-61601 2
Test Leads (red), Alligator clip to BNC(m) 16060-61602 2
Operation and Service Manual 16060-90000 1