User`s manual
2-Frame System - Command Module with External
Computer
NOTE A system using a Command Module as the resource manager identifies
instruments by an instrument identifier address. Instrument identifier
addresses are those divisible by 8 and must be used if you are using an
Agilent E1406A Command Module as the resource manager.
• MAINFRAME 1: Set the VXI-MXI module to address 2. Set other
VXI modules in the mainframe to addresses below 128. Note: The
resource manager must be at address 0.
• MAINFRAME 2: Set the VXI-MXI module address to address
128.
INTX
Cable
Agilent E1482
Resource Manager/ Slot
0 Device
GPIB Cable
Agilent E1406A
External
Computer
MXIbus Cable
Mainframe 2
Mainframe 1
Agilent E1482
VXI-MXI Module
LADDR = 128
Other Modules
LADDRs 129 - 255
Command Module
LADDR = 0
VXI-MXI Module
LADDR = 2
Other Modules
LADDRs 3 - 127
Figure 3-12. Command Module System, 2-Frame
3-12 Set Up the System for Multiple Mainframes C-Size Configuration Guide