User`s manual
MXIbus-3 : Remove the Terminating Networks from Middle VXI-MXI
Modules
If you have more than two mainframes, remove the MXI and INTX
terminating networks from the VXI-MXI modules in the middle of the
MXIbus daisy chain.
The VXI-MXI is shipped from the factory with terminating SIP resistor
networks installed. If the VXI-MXI will be the first or last device in the
MXIbus daisy-chain, you should leave these internal terminators in place.
If the VXI-MXI is not going to be an end device, the terminating resistor
networks must be removed from their sockets and stored in a safe place in
case the MXIbus system changes. All six MXIbus networks must be either
installed or removed.
Like the MXIbus, the INTX cables have matched impedance and require
termination networks at the first and last devices in the INTX chain. These
terminations minimize the reflections caused by impedance discontinuities
at the ends of the cable. The INTX daughter board comes with terminating
resistors installed. If the daughter board is not going to be an end device,
remove the terminating resistor network and store them with the MXI
networks.
MXI
Cable
INTX
Cable
First VXI-MXI in Chain: MXI
and INTX Terminating
Networks Present
Last VXI-MXI in Chain: MXI
and INTX Terminating
Networks Present
REMOVE MXI and INTX
Terminating Networks
INTX
INTX
INTX
MXI
MXI
MXI
Figure 3-10. MXIbus Daisy Chain