User`s manual
MXIbus-2 : Set VXI-MXI Modules for Non-Slot 0
VXI-MXI modules installed in a root mainframe (mainframe 1) that does
not contain an Agilent RADI-EPC7 or an external computer with an
EISA/ISA to MXIbus interface card, need to be configured for non-slot 0
operation. When setting a VXI-MXI module for non-slot 0 operation be
sure the module is configured as indicated in the center column in Table
3-1. The switch/jumper positions are shown on the next three pages.
Remember, if you have an EPC7 embedded computer or an external
computer with an EISA/ISA to MXIbus interface card, the VXI-MXI
module needs to be configured for slot 0 operation (MXIbus-1 : Set
VXI-MXI Module for Slot 0 Operation, this is the default configuration).
NOTE When installing the VXI-MXI module into a slot other than slot 0, you must
change switches S1 and S8 to configure for non-slot 0 use or damage can
occur.
Figure 3-3. Non-Slot 0 VXI-MXI Settings
C-Size Configuration Guide Set Up the System for Multiple Mainframes 3-5