EPC7-4B : Set the Command Module’s Servant Area
Notice the following when setting the Command Module servant area:
• For the Command Module to be the commander of a register-based
module, the register-based module’s logical address must fall within
the Command Module’s servant area. The servant area of the
Command Module is determined as:
Servant area = (logical address + 1) through (logical address +
servant area switch setting)
• The logical address plus the Command Module’s servant area cannot
exceed 255. Therefore, set the servant area based on the logical
addresses of the register-based modules in your system. For
example, if the Command Module’s logical address is 1 and its
servant area switch is set to 100, the Command Module would be the
commander for all modules with logical address from 2 through 101.
See Appendix A, "Terms and Definitions," for more information on the
servant area.
Servant Area Switch
E1406A
The servant area is the decimal sum of the switches in the
closed (1) position
Figure 2-44. Setting the E1406 Servant Area
C-Size Configuration Guide Setting Up an Embedded RADI-EPC7 Computer 2-57