User`s manual
EPC7-2 : Install the EPC7 into the Mainframe
Use the following procedure to install the Agilent RADI-EPC7 Embedded
Computer into slots 0 and 1 of the mainframe.
NOTE If you are using multiple mainframes via MXIbus, install the VXI-MXI
module into slot 0 (according to "Procedure 3 : Set Up Your System for
Multiple Mainframes") and the EPC7 into slots 1 and 2.
1. Turn off mainframe power by pressing the button in the lower left
corner of the mainframe.
2. Insert the module into the mainframe by aligning the top and
bottom of the EPC7 with the card guides inside the mainframe.
Slowly push the EPC7 straight into the slots until it seats in the
backplane receptacles. The front panel of the EPC7 should be even
with the front panel of the mainframe.
3. Tighten the retaining screws on the top and bottom edges of the
front panel.
NOTE The EPC7 has a front panel key that prevents its insertion next to certain
other types of modules. This is done to prevent problems associated with
incompatible signal levels on the VXI daisy-chained Local Bus. See the
Agilent RADI-EPC7 documentation for more information.
2-52 Setting Up an Embedded RADI-EPC7 Computer C-Size Configuration Guide