User`s manual

V743-2D : Disabling
the Command
Module’s Slot 0
Capability
Because the V743 controller contains the resource manager and slot 0
functionality that is also provided by the Command Module, the slot 0 and
VME system controller capability of the Command Module must be
disabled. This is done by setting the "System Controller" and "Slot 0"
switches to the "Disable" position. The location of these switches is shown
in Figure 2-35.
Note When using the Agilent E1406A Command Module with the V743
controller, an interrupt line other than line 1 is assigned to the Command
Module. In order for the Command Module and the instruments in its
servant area to function properly, the devices must use the same interrupt
line. Refer to "Assigning Interrupt Lines" in step "V743-4B: Apply Power"
for more information.
Figure 2-35. Disabling the E1406A Slot 0 Capability
C-Size Configuration Guide Setting Up an Embedded V743 Controller 2-41