User`s manual

Series 700 computer’s (GPIB) interface select code (typically 7), to form
the Command Module’s complete GPIB address.
Series 700-2D :
Disabling the
Command Module’s
Slot 0 Capability
Because the Agilent E1482B VXIbus Extender module is the system’s slot
0 module, the slot 0 and VME system controller capability of the Command
Module must be disabled. This is done by setting the "System Controller"
and "Slot 0" switches to the "Disable" position. The location of these
switches is shown in Figure 2-18.
Series 700-2E :
Disabling the
Command Module’s
VMEbus Time Out
Capability
When E1482 VXIbus Extender modules are part of your system, the
modules must function as the Data Transfer Bus (DTB) timer. This means
the bus timer capability of the Command Module must be disabled. This is
done by setting the ’VME BTO Disableswitch as indicated in Figure 2-19.
The VMEbus timer capability (VME BTO Level) of the E1482B is enabled
at the factory.
Figure 2-18. Disabling the E1406A Slot 0 Capability
2-26 Setting Up a Series 700 Computer C-Size Configuration Guide