User`s manual

Select the Add button.
Repeat this procedure for each VXI instrument in your system.
3. Select OK to end.
Assigning Commander/Servant Hierarchy
1. From the VXI Configurator, select VXI Control and scroll to
Commander hierarchy.
2. Select the commander (Cmdmod, for example) from the All known
names box and select the Add commander button. The
commander added will now be shown in the Commanders box.
3. Now, select the commander (Cmdmod, for example) from the
Commanders box and select the device you want to add as a
servant in the All known names box. Select the Add servant
button to add the device as a servant to the commander. (The
devices listed in the All known names box are the same devices
you assigned names in the "Assigning VXI Device Names" section.)
Repeat this step for each device to add as a servant to the
commander.
4. Select OK to end.
Assigning Interrupt Lines
1. From the VXI Configurator, select VXI Control and scroll to
Interrupt mapping.
2. Select the Commander (Cmdmod, for example) from the Selected
device box.
3. Clear the Assign IRQs at runtime box so that you can assign
interrupt lines to the selected commander. Assign the interrupt
lines as follows:
In the attributes box, assign interrupt line 1 to the commander by
placing a 1 under H 1 in the Handlers, IRQ number or 0 box.
Select the Change button to change interrupt line assignments.
4. If you now have two commanders assigned to interrupt line 1, select
the other commander and change H1 in the Handlers, IRQ
number or 0 box so that interrupt line 1 is assigned only to the
Command Module.
5. Select OK to end.
5-12 Applying Power C-Size Configuration Guide