User`s manual
• if necessary, call your nearest Agilent Technologies sales and service
office.
When using the Command Module for the first time or when the mainframe
has not been turned on for at least one week, leave the mainframe on for at
least 15 hours to fully charge the Command Module’s battery.
If a configuration or start-up error such as an invalid address or failed self
test occurs, the error is reported in the power-on and configuration
sequence. A list of the configuration and start-up error messages and their
causes can be found at the end of this guide.
HP 9000 Series 700
Computer
To verify that the E1482B VXIbus extender module, the MXIbus and INTX
cables, and all VXI instruments are correctly installed, turn on the
mainframe. This starts the system resource manager function (ivxirm)
provided by the Agilent E1489I MXIbus Controller Interface card and the
Agilent E2093B SICL software. To view the contents of the configuration
file (rsrcmgr.out) written to by the resource manager, type the following
command:
/usr/pil/bin/ivxisc
An example of a configuration file (sequence) is shown in Figure 5-2.
5-4 Applying Power C-Size Configuration Guide