Step-2 : Set Instrument Logical Addresses
All instruments are shipped with a factory-set logical address. In some
cases you can use the factory-set logical address. The following table lists
selected Agilent instruments and their factory-set logical addresses:
Notice that some cards have the same logical address. When installing
cards into a mainframe, however, each card must have a unique logical
address. Thus, if you have cards with the same factory-set logical address,
you must change the logical address on some of the cards.
Use the following steps to set an instrument’s logical address:
1. Locate the logical address switch on the card. Figure 4-1 shows the
switch locations for most B- and C-Size modules.
2. Set the card’s logical address using the following rules:
First, try to use the factory-set logical address. If you have
modules with the same factory-set logical address, change the
address of the modules until all modules have different logical
addresses.
If you have multiple mainframes connected via MXIbus, make
sure the logical addresses of the modules in a mainframe are
within the logical address window for that mainframe. See
"MXIbus-4 : Set the VXI-MXI Module’s Logical Address" in
procedure 3.
Agilent Instruments Logical
Address
Agilent Instrument Logical
Address
Digital Multimeters
Agilent E1326B/E1411B
24 Multiplexers
Agilent E13xx Multiplexers
Agilent E14xx Multiplexers
112
Digitizers
Agilent E1426A, E1428A
Agilent E1429A/B
40 Other Switches
Agilent E13xx, E14xx Matrix
Agilent E13xx, E14xx RF
Agilent E13xx, E14xx Form C
Agilent E13xx, E14xx General Purpose
120
Counters
Agilent E1420B, E1332A, E1333A
48 Digital Functions Test
Agilent Model D20
136
Power Meters
Agilent E1416A
56 Digital Input / Output
Agilent E1330B
144
Arbitrary Waveform Generators
Agilent E1340A, E1445A
80 Agilent Instrument BASIC 240
Table 4-2. Factory-Set Logical Addresses
4-4 Configure and Install Instruments C-Size Configuration Guide