User`s manual

MXIbus-5 : Disable the VMEbus Timeout on Other Modules (VME BTO)
The VXI-MXI module must do the VMEbus timeout on all mainframes.
You MUST disable the VME BTO on any other device in your system. The
VME BTO must be disabled on the following:
Agilent E1406A Command Module
Agilent RADI-EPC7 Embedded Computer
NOTE You must disable the VMEbus timeout on any VXI module used in the
VXI-MXI system. Any module with the VME BTO enabled will not allow the
system to be configurable. Only the VXI-MXI module in each mainframe
can have the VME BTO enabled.You do not have to disable the VMEbus
timeout on the Agilent V743/64 and V743/100 embedded controllers.
Disable VMEbus Timeout - Agilent E1406A
Command Module
Agilent E1406A
VME BTO Shown
Disabled
VME BTO Shown Disabled
OPEN = Switch Set to 0 (OFF)
CLOSED = Switch Set to 1 (ON)
Figure 3-20. Disabling VMEbus Timeout, Command Module
3-20 Set Up the System for Multiple Mainframes C-Size Configuration Guide