User`s manual

3-Frame System - Agilent V743 Embedded
Controller
MAINFRAME 1: Set the VXI-MXI module to address 2. It does
not have to be an instrument identifier address. Set other VXI
modules in the mainframe to addresses below 128. Note: the
resource manager must be at address 0.
MAINFRAME 2: Set the VXI-MXI module address to address
128. Set all other VXI modules to addresses between 128 and 191
(do not duplicate the VXI-MXI address).
MAINFRAME 3: Set the VXI-MXI module address to 192. Set all
other VXI modules to addresses between 192 and 255 (do not
duplicate the VXI-MXI address).
VXI-MXI Module
LADDR = 128
Other Modules
LADDR’s 129 - 191
Embedded Controller
LADDR = 0
VXI-MXI Module
LADDR = 2
Other Modules
LADDR’s 3 - 127
Agilent E1482
Agilent E1482
Mainframe 1
Mainframe 3
Embedded Controller
Logical Address = 0
MXIbus Cable
INTX Cable
VXI-MXI Module
LADDR = 192
Other Modules
LADDR’s 193 - 255
Agilent E1482
Mainframe 2
Figure 3-19. Embedded V743, 3-Frame
C-Size Configuration Guide Set Up the System for Multiple Mainframes 3-19