Specifications

7
Backplane Specifications
Solid-state automatic daisy-chain jumpering for BUS
GRANT and IACK signals.
Full differential distribution of CLK10.
ACFAIL* and SYSRESET* in full compliance with the
VMEbus and VXIbus Specifications.
Surface mount construction and no sockets for maximum
reliability.
Basic Monitor Specifications
(Applicable to all three mainframes)
Indicators: • Power-on or Standby status
• Power supply output voltages
monitor
• Power supply temperature monitor
• Fans status monitor
• Backplane activity monitor
• Backplane SYSFAIL monitor
Switches: • On/Standby
• Fan Mode: Switches all fans
between Full Speed and Variable
Speed modes
• Reset: Asserts backplane signals
SYSRESET and ACFAIL
Diagnostic connector: • Output all 7 backplane voltages
for monitoring
• Output +5 V and +12 V for remote
applications. 1A max each
• Input +5VSTDBY to backplane. 1A
max total for pins 5 and 18
• Remotely operate On/Standby
• Power supply temperature output
• Reference temperature output
• Fans OK output, same as Fans
indicator
• Backplane voltages OK output
• SYSRESET*, input or output
• ACFAIL*, output
• Ground
All other slots blocked. Airflow decreases as additional
slots are opened.
Performance shown for Worst Slot (slot 2). Airflow is
greater in all other slots.
Fans on Full Speed. Minimum airflow is approximately
50% with fans on Variable Speed.
Air Filter Kit not installed. Airflow is reduced
approximately 10% with clean air filters installed.
Measurements taken at 1,500 m altitude.
Environmental
Temperature
Operating temperature range: 0° C – +55° C
Storage temperature range: –40° C – +75° C
Humidity
Operating humidity range: Up to 95% RH from 0° C to +40° C
Up to 65% RH from +40° C to +55° C
Storage humidity range: Up to 95% RH from 0° C to +55° C
Up to 65% RH from +55° C to +75° C
Acoustic Noise
(Sound power at bystander position one meter in front of mainframe)
High-speed fan: 55 dBA
Variable fan on low speed: 42 dBA
Shielding
Front panel EMC gasketing: Front panel gasketing provided per
VXI
Rev. 1.4, B.7.2.3
Backplane shielding: Backplane connector shields per VXI
Rev. 1.4, B.7.2.3
Intermodule chassis shielding: Intermodule chassis shields per VXI
Rev. 1.4, B.7.3.4
Altitude: Up to 3000 m
(Diagnosis and troubleshooting through the front panel monitor and
connector.)
MTTR = Mean Time to Repair
MTTR, power supply: <10 min. (w/mainframe and modules
fully installed in rack)
MTTR, impeller and/or fan: <10 min. (w/mainframe and modules
fully installed in rack)
MTTR, E8404A enhanced monitor
control board:
<5 - 10 min. (w/mainframe and
modules fully installed in rack)
Repair
100% compatible with the VXIbus Specification
Revision 1.4
• E8404A command set compatible with IEEE-488.1,
IEEE-488.2, and SCPI-1995.0
Standards Compliance